microAIR Alternating Pressure Mattress with 50 LPM Compressor Inavcare MA60
Economical alternating pressure mattress replacement system. This Group II support surface uses 20 - 8" Bladders and has a 1000-lb weight capacity.
Alternating pressure therapeutics support system with 50 LPM compressor, automatic pressure monitoring and automatic wireless fowler.
The microAIR APM provides pressure relief in the prevention and treatment of pressure ulcers.
Designed for the treatment of pressure ulcers, this therapeutic support system delivers alternating pressure with a powerful 50 LPM compressor. Product will not bottom out as long as mattress hose stays connected to the compressor even in a power outage. Wireless fowler automatically adds 35% more air when bed is articulated.
microAIR Technology:
New
wireless fowler technology, real time pressure monitoring and energy
conserving power units combine to make microAIR surfaces the most
efficient dynamic support surfaces in the market. Within the power
units is a pressure transducer that maintains surface pressure not to
exceed 32mm/Hg. The solenoid valve within the compressor opens and
closes the air valve automatically to maintain the pressure relief
range prescribed. The compressor turns off to save energy when the
patient is in the range.
Wireless Fowler Technology is an
innovation exclusive that assures patient comfort. From time to time
patients need to sit up in bed, which creates the problem of additional
pressure in the coccyx area and possible disruption of treatment.
Working
in sync with the power unit, the transmitter installed in the head
section of the mattress signals the power unit when the head portion
begins to raise. The power unit immediately begins adding 25% more air
to the mattress to prevent patient contact with a hard surface.
microAIR Control Panel:

microAIR Control Panel. Treatment times are 5, 10, 15, and 20 minute cycles.
Alternating Pressure Mode: The alternating pressure times indicate the frequency of deflation or inflation of half of the air cushions (even or odd numbered). For example, 10 minutes is selected. In this example, the even numbered air cushions in the mattress will deflate, while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed, the even numbered air cushions will inflate while the odd numbered cushions remain at constant pressure. When 10 minutes has elapsed again, the odd numbered air cushions deflate while the even numbered cushions remain at constant pressure. The air cushions continue to inflate or deflate like this as long as an alternating pressure time has been selected. At all times, half of the air cushions remain at a constant pressure when an alternating pressure time has been selected.
Static Mode: all air cushions in the mattress are maintained at constant pressure. The Static LED is illuminated GREEN.
Minimum Pressure Setting: The minimum pressure setting during Alternating Pressure is set to 0 (2 mmHg) at the factory. This means that the mattress will deflate as much as possible during the deflation time when using 10, 15 or 20 minute Alternating Pressure times. The setting can be changed so the mattress will only deflate to 50% of the selected pressure setting.
microAIR Mattress:
The waterproof cover is made of non-shear nylon and has nylon tie down straps with buckles for attaching the mattress to the bed. There is also a 2" foam safety pad under the nylon cover.
This microAIR mattress has a non-skid bottom and is secured with 10
nylon web buckled hold-downs. When rolled up for storage, carrying
handles allow for easy transport.
Standard Hospital/Homecare Bed Size: Affordable with added comfort, this alternating pressure mattress is available in the standard 36" x 80" size (Twin XL). |
|
 |
 |
|
 |
 |
 |
 |
 |
Features:
- Dual quilted cover with slip layer helps prevent friction and shear.
- Automatic wireless fowler prevents patient from bottoming out
- 50 LPM compressor is easily set by patient weight and remains in pressure relief range
- Alternating pressure cucles are 5, 10, 15 and 20 minutes
- 20 horizontal bladders deliver alternating pressure
- 20 - 8" Bladders
- Dimensions: 36" x 80" x 8"
Specifications:
| Flammability: |
16 CFR 1632 and 16 CFR 1633
compliant
|
| Overall Height: |
10"
|
| Overall Length: |
80"
|
| Overall Width: |
36"
|
| Product Weight
Capacity: |
1000 lbs
|
| Typical Use: |
Prevention and relief of pressure
ulcers
|
| Warranty: |
2 years compressor, 1 year on
mattress
|
|
|
 |
 |
 |
 |
|
 |

|